Prostatic thermoluminescent dosimeter analysis in a patient treated with 18 MV X rays through a prosthetic hip.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 8420884.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
External beam radiation therapy with high energy photon beams through hip prostheses has been shown to cause dose inhomogeneities for target volumes in the pelvis. In this work, measurements of dose using thermoluminescent dosimetry were compared with dose calculations from a computerized treatment planning system in a patient with prostatic carcinoma and a cobalt-chromium-molybdenum hip prosthesis. A 39% decrement in dose at isocenter was demonstrated for an 18 MV photon beam passing through the prosthesis. A discrepancy of only 3.1% was shown between measured and calculated dose when the tissue-maximum ratio (TMR) method of heterogeneity correction was used. However, it is recognized that several sources of error are possible when heterogeneity corrections are performed for high density prostheses and these are discussed below. The results of this work stress the importance of accurate data for use with the "ratio of TMR's" algorithm in order that accurate treatment planning can be performed.
Medical subject headings
- Hip Prosthesis
- Prostatic Neoplasms