Development of a device for measuring adherence of skin grafts to the wound surface.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 8434820.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Adherence of a biological graft to the wound surface is the most important factor influencing the ultimate success of graft viability. A machine has been developed to test the adherence of biological graft materials to a substrate such as a wound surface. The peeling mode, which yields reproducible quantitative measurements of adherence, is a standard method for testing adhesives. The device is designed to continuously measure the force required to peel the graft from the substrate at a constant rate. This force is a function of the energy of adhesion per unit area of adhered surface. This device has been used to measure the peeling force of (2 x 2 cm) skin grafts which are applied to full-thickness wounds on mice. Results of tests on adherence of autografts on mice show that the peeling force increases significantly with time over the first 9 days of healing. Thus, this device is useful in quantitative comparison of various skin grafting techniques and artificial grafts.
Medical subject headings
- Materials Testing
- Skin Transplantation
- Wounds and Injuries