Pull-out strength of five suture anchors.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 8442817.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Five commercially available suture anchor devices were tested to failure in human cadaveric proximal tibiae. A total of 198 trials were completed. Insertion was uncomplicated for all anchors, with the exception of the Acufex Rod TAG device, with five of 22 (23%) of these breaking upon insertion into thicker cortical bone. Overall, the anchors performed significantly (p < 0.05) better when placed in thicker cortical bone, further away from the joint surface, and when loaded in a direction parallel to the bone surface. The Mitek GII anchor failed (pulled out of bone) less often than the other anchors (19% vs. 46%). In loading perpendicular to the bone surface, the Mitek GII and Statak were the strongest (p < 0.05), with mean loads at failure of 82.5 and 90.2 N, respectively. The Acufex Rod TAG, Acufex Wedge TAG, and Mitek GI failed at 67.2, 65.5, and 49.4 N respectively.