MR imaging of silicone breast implants: comparison of different coil arrays.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 8451430.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To determine the optimal coil and patient position for magnetic resonance imaging of patients with silicone breast implants, images obtained with body, circular surface, flexible, single shoulder, and dual shoulder coils in various numbers of patients were compared with regard to signal-to-noise ratio (S/N), decay of signal intensity, resolution, and key area coverage. Images obtained with the dual shoulder coil and prone position, with a customized positioning device, gave the best S/N, resolution, and coverage of the breasts and axillae.
Medical subject headings
- Breast
- Magnetic Resonance Imaging
- Prostheses and Implants
- Silicones