TiN coating: surface characterization and haemocompatibility.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 8476989.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The left ventricular assist device under consideration is based on the principle of the Maillard-Wankel rotary pump. The construction materials must meet stringent requirements. Titanium nitride was chosen for its surface properties and graphite for its bulk characteristics. The purpose of this study was to characterize the chemical vapour deposition titanium nitride coating via morphology, roughness, crystallinity, chemical composition, to report and discuss the results of in vitro haemocompatibility tests (protein adsorption, platelet retention, haemolysis) and to discuss physico-chemical and biological results. This chemical vapour deposition titanium nitride coating is well tolerated by the blood despite its surface irregularities, and appears as a good candidate material after improvements.
Medical subject headings
- Heart-Assist Devices
- Titanium