Radiation dose to positron emission tomography technologists during quantitative versus qualitative studies.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 8478709.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Positron emission tomography technologists were monitored with thermoluminescent dosimeters (TLDs) during qualitative and quantitative studies. Doses to technologists during specific tasks were also measured. The technologists received at least twice as much radiation during the quantitative as the qualitative studies. The average dose per study for qualitative studies was 0.017 mSv (1.7 mrem) shallow and 0.014 mSv (1.4 mrem) deep. The average dose per study for the quantitative studies was 0.05 mSv (5 mrem) shallow and 0.04 mSv (4 mrem) deep. The average dose per study was based on the TLD dose accumulated over studies conducted over four 2-mo and one 1-mo intervals. The dose incurred by the technologists each time they drew a radioactive dose was 0.002 mSv (0.2 mrem) shallow and 0.001 mSv (0.1 mrem) deep. The doses received during injection were 0.014 mSv (1.4 mrem) shallow and 0.007 mSv (0.7 mrem) deep. Doses received during blood sampling were 0.016 mSv (1.6 mrem) shallow and 0.014 mSv (1.4 mrem) deep. During quantitative studies, the technologist received a much greater dose than during its qualitative counterpart due to the blood sampling process and increased time in the room with the radioactive patient.
Medical subject headings
- Occupational Exposure
- Radiation Monitoring
- Technology, Radiologic
- Tomography, Emission-Computed