Radiation dose to positron emission tomography technologists during quantitative versus qualitative studies.

McCormick, V A; Miklos, J A · J Nucl Med · 1993

retrospective_cohort · Level III

Where this comes from

Abstract

Positron emission tomography technologists were monitored with thermoluminescent dosimeters (TLDs) during qualitative and quantitative studies. Doses to technologists during specific tasks were also measured. The technologists received at least twice as much radiation during the quantitative as the qualitative studies. The average dose per study for qualitative studies was 0.017 mSv (1.7 mrem) shallow and 0.014 mSv (1.4 mrem) deep. The average dose per study for the quantitative studies was 0.05 mSv (5 mrem) shallow and 0.04 mSv (4 mrem) deep. The average dose per study was based on the TLD dose accumulated over studies conducted over four 2-mo and one 1-mo intervals. The dose incurred by the technologists each time they drew a radioactive dose was 0.002 mSv (0.2 mrem) shallow and 0.001 mSv (0.1 mrem) deep. The doses received during injection were 0.014 mSv (1.4 mrem) shallow and 0.007 mSv (0.7 mrem) deep. Doses received during blood sampling were 0.016 mSv (1.6 mrem) shallow and 0.014 mSv (1.4 mrem) deep. During quantitative studies, the technologist received a much greater dose than during its qualitative counterpart due to the blood sampling process and increased time in the room with the radioactive patient.