Factors involved in catheter obstruction during long-term peritoneal insulin infusion.
case_series · Level IV
Where this comes from
- Record sourced from PubMed, PMID 8495622.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To analyze the efficacy of ECPII and the factors responsible for technical problems often encountered. This treatment has been in use with IDDM patients since 1980. Forty-four IDDM patients were treated by ECPII for 42-78 mo (mean, 53 mo). Glycemic equilibrium was improved during treatment (mean plasma glucose level, 7.6 mM; mean GHb level, 8%). Catheter blockage was the main reason for ECPII failure (74%). Mean catheter survival of each catheter, determined by actuarial analysis, was 11.7 mo and significantly decreased with subsequent implantation. SEM of the catheter tips showed deposits composed of fibrin and cells occluding the inner lumen. Factors such as age, sex, local infection, and low insulin basal rate were not found to have any incidence on the catheter survival. Placement of the catheter in the upper part of the peritoneum, however, increased catheter survival. Anti-insulin antibodies did not seem to be directly involved in blockage. We conclude from this long-term experience that during ECPII, catheter blockage remains the major recurring complication, probably involving a local immune-inflammatory response in the peritoneum.
Medical subject headings
- Catheters, Indwelling
- Diabetes Mellitus, Type 1
- Insulin Infusion Systems