Fresh gas requirements of an enclosed afferent reservoir breathing system in anaesthetized, spontaneously breathing adults.
other
Where this comes from
- Record sourced from PubMed, PMID 8499214.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Using two methods of determining the onset of rebreathing, we have determined the minimum fresh gas flow rate (VF) of the Ohmeda enclosed afferent reservoir breathing system (EAR) in anaesthetized, spontaneously breathing adults. Rebreathing as defined by the Kain and Nunn criteria did not occur when the VF/VE ratio was greater than 0.70. A mathematical model was used to calculate the degree of rebreathing at each VF. From this model, rebreathing did not occur when VF was 0.86 VE or more and this value of VF/VE is considered appropriate to eliminate rebreathing in clinical practice.
Medical subject headings
- Anesthesia, General
- Ventilators, Mechanical