Negative Poisson ratios and strain-dependent mechanical properties in arterial prostheses.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 8519933.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The elastic properties of two artificial arterial prostheses have been measured. One is a fibrillar polyurethane (PU) material, the other is an expanded polytetrafluoroethylene (ex-PTFE). It is shown that the conventional assumption of incompressibility is invalid for both materials and that changes in the thickness of the prosthetic artery wall must be measured directly. The ex-PTFE shows highly strain-dependent properties, the most significant being that of large negative through-thickness Poisson ratios, up to -11. These results have important consequences for matching the properties of prostheses to natural materials.
Medical subject headings
- Blood Vessel Prosthesis
- Polytetrafluoroethylene
- Polyurethanes