Assessment of energy-weighted acquisition in SPECT using ROC analysis.
other
Where this comes from
- Record sourced from PubMed, PMID 8523131.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Energy-weighted acquisition is a technique for reducing Compton scatter effects in nuclear medicine images. The effect of energy-weighted acquisition on SPECT99mTc images was evaluated by comparing the energy-weighted acquisition images with those obtained using a 20% photopeak energy window. SPECT images were compared using receiver operator characteristic (ROC) experiments testing the observer's ability to perform a pseudoclinical task. The tasks were detecting cold lesions within a uniform background and cold lesions within images created with a tomographic brain phantom. ROC analysis for each phantom produced different results. No significant difference was found between the two acquisition techniques in detecting cold lesions on uniform backgrounds. Energy-weighted acquisition improved cold lesion detection significantly within the brain phantom in comparison with 20% photopeak acquisition. Lesion detection in 99mTcSPECT images can be improved using energy-weighted acquisition. This improvement, however, is dependent on the nature of the object being imaged. Images with structure show improved detection, whereas uniform images do not.
Medical subject headings
- Brain
- Image Processing, Computer-Assisted
- Technetium
- Tomography, Emission-Computed, Single-Photon