Pressure gradients across bileaflet aortic valves by direct measurement and echocardiography.

Laske, A; Jenni, R; Maloigne, M; Vassalli, G; Bertel, O; Turina, M I · Ann Thorac Surg · 1996

retrospective_cohort · Level III

Where this comes from

Abstract

Pressure gradients calculated from echo-cardiography after aortic valve replacement are commonly much higher than would be expected from in vitro measurements. The mean, peak-to-peak, and maximal gradients across bileaflet aortic prostheses (St. Jude Medical) were measured invasively in 52 patients at high and low heart rate, cardiac index, and stroke volume. One week after operation the gradients were calculated from a standard transthoracic echocardiogram (delta p = 4v2(2)). In a second study 3 to 12 months later, gradients were calculated using the standard, simplified Bernoulli equation, and with the equation considering subvalvular flow velocities (delta p = 4(v2(2-)v1(2)). Invasive and echocardiographic measurements were matched and compared. Invasively measured mean gradients for 21 to 29-mm valves ranged from 7.4 +/- 4.9 to 4.3 +/- 1.6 mm Hg at systolic flow rates from 11.3 +/- 0.7 to 16.2 +/- 1.8 L.min-1.m-2. Mean echocardiographic gradients were 15.1 +/- 4.5 to 7.5 +/- 2.2 mm Hg (p < 0.001) with the standard method, and 10.5 +/- 1.9 to 5.6 +/- 1.5 mm Hg when considering the subvalvular flow velocity (p < 0.001). Mean gradients across bileaflet prostheses are generally low, even in small valves and with high systolic flow. The correlation of the invasive in vivo with in vitro gradients is good. Standard echocardiography overestimates gradients across bileaflet heart valves and high gradients are not due to valve dysfunction. Gradients obtained by echocardiography considering the subvalvular flow velocity correlate better to invasively measured and in vitro gradients.

Medical subject headings