Monitoring of mixed venous oxygen saturation and pressure from biosensors in the right ventricle. A 24 hour study in patients with heart failure.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 8582384.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Right ventricle venous oxygen saturation was studied in 10 heart failure patients over 24 h using a lead equipped with an oxygen biosensor inserted temporarily. At the same time a pressure sensor, incorporated in the same lead to measure right ventricular pulse pressure and maximum positive and negative rate of pressure development (RV dP/dt), was tested to see whether it would interfere with the oxygen sensor. Data from the biosensor lead were continuously compared with mixed venous oxygen saturation obtained from a fibreoptic Swan-Ganz catheter with the tip in the pulmonary artery. For reference, blood samples were drawn at regular intervals from this catheter. A provocative protocol was used to cause haemodynamic changes. There was good correlation between oxygen saturation in the right ventricle (biosensor-derived) and mixed venous oxygen saturation in the pulmonary artery (fibreoptic-derived) (r = 0.86) and between sensor-obtained right ventricular oxygen saturation and oxygen saturation in the blood samples from the pulmonary artery (r = 0.90). Changes in central haemodynamics were also well reflected by changes in pulse pressure and dP/dt derived from the pressure sensor. As it was not possible to obtain absolute pressures no attempt was made to compare the pressure changes with conventional haemodynamics. These 24 h haemodynamic measurements suggest that a biosensor-equipped lead placed in the right ventricle could be a valuable tool for long-term monitoring of mixed venous oxygen saturation and pressure in patients with congestive heart failure.
Medical subject headings
- Biosensing Techniques
- Heart Failure
- Monitoring, Physiologic
- Oxygen
- Ventricular Pressure