Reduction of motion artifacts using a two-frequency impedance plethysmograph and adaptive filtering.
Where this comes from
- Record sourced from PubMed, PMID 8582723.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We measured transthoracic impedance in nine presumed healthy adult subjects with a two-frequency plethysmograph at 57 kHz and 185 kHz. The measurement protocol included periods of normal breathing without motion and periods of motion without breathing. We analyzed the cross-correlation and the ratio between the signals at both frequencies for all the different maneuvers. The correlation coefficient was between 0.97 and 1 for breathing, the minimal cross-correlation (0.81) was for simulated obstructive apnea. We found that the amplitude ratio between the two-frequency signals was different for normal breathing and for motion. Based on these results, we designed and tested an adaptive filter to increase the signal-to-artifact ratio (SAR). The increase in SAR (mean +/- standard deviation) compared with the signal at 57 kHz was: 183% +/- 117% for arm movement, 133% +/- 93% for leg movement, and 34% +/- 62% for simulated obstructive apnea.
Medical subject headings
- Artifacts
- Plethysmography, Impedance