Development of an integrated radiotherapy network system.

Nagata, Y; Okajima, K; Murata, R; Mitsumori, M; Mizowaki, T; Yamamoto, M; Hiraoka, M; Nishidai, T et al. · Int J Radiat Oncol Biol Phys · 1996

Where this comes from

Abstract

To introduce the process of developing an integrated radiotherapy network. We developed a new radiotherapy treatment-planning system in 1987 that we named the Computer Tomography (CT) simulator. CT images were immediately transported to multiimage monitors and to a planning computer, and treatment planning could be performed with the patient lying on the CT couch. The results of planning were used to guide a laser projector, and radiation fields were projected onto the skin of the patient. Since 1991, an integrated radiotherapy network system has been developed, which consists of a picture archiving and communicating system (PACS), a radiotherapy information database, a CT simulator, and a linear accelerator with a multileaf collimator. Clinical experience has been accumulated in more than 1,000 patients. Based on our 7 years of experience, we have modified several components of our original CT simulator and have developed a second generation CT simulator. A standard protocol has been developed for communication between the CT scanner, treatment planning computer, and radiotherapy apparatus using the Ethernet network. As a result, treatment planning data can be transported to the linear accelerator within 1 min after completion of treatment planning. This system enables us to make optimal use of CT information and to devise accurate three-dimensional (3D) treatment-planning programs. Our network also allows for the performance of fully computer-controlled dynamic arc conformal therapy.

Medical subject headings