Right ventricular blood flow during left ventricular support in an experimental porcine model.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 8607683.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Right ventricular blood flow may be adversely affected during left ventricular assist device (LVAD) use leading to right ventricular (RV) ischemia and RV dysfunction. This study characterized normal RV blood flow responses to LVAD operation. Seven Yorkshire pigs weighing 74.4 +/- 3.4 kg underwent right coronary artery blood flow measurements with an ultrasonic flow probe and injection of radiolabeled microspheres. A Thoratec LVAD was used in either synchronous or asynchronous modes and RV loading was increased using a pulmonary artery snare. The RV blood flow was compared between three regions that differed in proximity to the right coronary artery: proximal segment, mid-RV, and distal. The right ventricular distal flow was 0.93 +/- 0.07 mL x min-1 x g-1 compared with 0.74 +/- 0.06 mL x min-1 x g-1 at right ventricular proximal flow during control measurements (p = 0.0001). This difference was maintained during LVAD operation in either synchronous or asynchronous modes and also during pulmonary artery constriction. Global RV flow is not adversely affected by LVAD use. A flow gradient occurs along the right coronary artery with the distal vascular bed having relatively less reserve, which may be more susceptible to ischemia in patients with preexisting coronary disease or RV distention during LVAD use.
Medical subject headings
- Heart-Assist Devices
- Models, Cardiovascular
- Ventricular Function, Right