Reversible clinical outcome after sphincter stent removal.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 8618306.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We determined whether the self-expanding sphincter stent, a potential alternative to conventional external sphincterotomy for the treatment of detrusor external sphincter dyssynergia, causes a permanent effect on the lower urinary tract. Four spinal cord injured men with voiding symptoms of detrusor external sphincter dyssynergia as noted by complete urological evaluation, including a video urodynamic study, were treated with the self-expanding sphincter stent. However, the device was explanted 6 months or longer after insertion in all 4 cases due to stent migration (3) and difficulty with condom catheter urinary drainage (1). All stents were removed completely without damage to the urethra. Mean voiding pressure decreased from 62.5 +/- 39.4 to 20.7 +/- 6.5 cm. water after sphincter stent placement. One year after stent explantation mean voiding pressure remained unchanged from preoperative values of 58.5 +/- 21.5 cm. water. No patient had stress urinary incontinence or endoscopically apparent urethral strictures. The stent can be removed even after complete epithelialization and an extended interval without damage to external sphincter function or urethral stricture formation. The urinary sphincter stent is an effective, reversible treatment for patients with detrusor external sphincter dyssynergia.
Medical subject headings
- Stents
- Urethra
- Urinary Bladder, Neurogenic