Component design affecting patellofemoral complications after total knee arthroplasty.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 8620639.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Three hundred one primary cemented total knee arthroplasties were performed in 289 patients. Two different prostheses were used; 148 knees received the Miller-Galante I prosthesis and 153 knees received the Press Fit Condylar prosthesis. Minimum followup was 2 years. The groups were similar in all parameters both preoperatively and postoperatively, with the exception of the patellofemoral complication rate. Knees that were implanted with the Miller-Galante I prosthesis experienced a complication rate of 10.1%, while those with the Press Fit Condylar prosthesis experienced a complication rate of 0.7%. The distinct difference in the patellofemoral complication rate may be due to the differences in design of the femoral component. Features that may have contributed to increased patellofemoral morbidity included a short, narrow anterior flange; a shallow patellar groove; and an abrupt anterior to distal transition with a smaller radius of curvature. Because subtle design difference can have a profound effect on clinical outcome, long term evaluation of new designs should be completed before widespread use. The clinician also should be aware of the desirable design features when choosing a component.
Medical subject headings
- Knee Prosthesis
- Postoperative Complications
Anatomy
- femur
- knee