Force, torque, and stress relaxation with direct laryngoscopy.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 8623942.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The anesthetist exerts axial force on the laryngoscope handle to expose the glottis. The anesthetist must also apply a perpendicular force to balance the torque on the laryngoscope. Several studies have measured axial force during direct laryngoscopy, but none has measured torque. This study used a newly designed laryngoscope handle to measure force and torque simultaneously during direct laryngoscopy of ASA grade I and II patients requiring general anesthesia and endotracheal intubation for elective surgery. In 58 patients, peak force averaged 38 +/- 2 newtons. Peak torque averaged 4 +/- 0.2 newton-meters, and the perpendicular force was estimated as 40 +/- 2 newtons. The peak torque that can be balanced by the wrist is approximately 6 newton-meters, suggesting that torque may be a limiting factor for laryngoscopy in some situations. Peak force and torque demonstrated stress relaxation, a viscous property of biologic tissues. Force and torque decreased monoexponentially to approximately 70% of peak values with a half-time of 4 +/- 0.3 s. The phenomenon occurred in spite of administration of muscle relaxants, and was probably due to stress relaxation of pharyngeal tissues that are passively stretched during laryngoscopy.
Medical subject headings
- Laryngoscopy