The importance of the expiratory pause. Comparison of the Mapleson A, C and D breathing systems using a lung model.
other
Where this comes from
- Record sourced from PubMed, PMID 8694159.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
A physical lung model simulating spontaneous respiration was used to investigate the influence of the respiratory pattern on the efficiency of the Mapleson A, C and D breathing systems. It is shown that the Mapleson A system is always the most efficient breathing system and that its performance is relatively independent of the respiratory pattern. When the expiratory pause is minimal, the Mapleson C system is almost as efficient as the Mapleson A, but becomes ever less efficient as the expiratory pause increases. The Mapleson D system is very inefficient when the expiratory pause is short. With a longer expiratory pause, this system's efficiency approaches that of the Mapleson A. The experimental results are compared with predictions generated by a mathematical model. There is good agreement between the two, validating the mathematics used.
Medical subject headings
- Anesthesia, Inhalation
- Lung
- Models, Anatomic
- Respiration