Overall accuracy of technetium-99m-MAG3 clearance measurements obtained with a gamma camera heart curve.

Tondeur, M; Kinthaert, J; Piepsz, A; Ham, H R · J Nucl Med · 1996

case_series · Level IV

Where this comes from

Abstract

The aim of the present study was to evaluate the accuracy of 99mTc-MAG3 clearance measurements using a precordial gamma camera curve calibrated by a single plasma sample. Technetium-99m-MAG3 was administered to ten young normal volunteers. A 60-min gamma camera acquisition was performed. Five different segments of the gamma camera curve were determined: 3 min to 20 min, 3 min to 30 min, 3 min to 40 min, 3 min to 50 min and 3 min to 60 min. A biexponential function was fitted on each of these five different segments, which were thereafter calibrated using eight different blood samples. These blood samples were successively used for calibration at 5, 10, 15, 20, 30, 40, 50 and 60 min. The single injection, multiple plasma sample method was used as reference. Camera clearances varied markedly based on the length of the precordial curve and on the time of the calibration sample. Different regression equations were obtained for each duration of the camera curve, and for each blood sample timing. Correlation coefficients were > 0.95 in most cases recording period of at least 50 min, however, was necessary to obtain a s.e.e. better than those obtained using a single blood sample method without gamma camera curve. The 99mTc-MAG3 clearance determination using a gamma camera heart curve calibrated with a single blood sample does not necessarily improve the accuracy of the one blood sample method.