Early failure of Hylamer acetabular inserts due to eccentric wear.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 8713919.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The authors performed a retrospective review of the survivorship of Hylamer (DePuy Dupont Orthopaedics, Warsaw, IN) acetabular liners following primary arthroplasty of the acetabulum. Between 1990 and 1992, 193 total hip arthroplasties were performed in which Hylamer acetabular liners were implanted. One hundred forty-three of these hips (74%) were available for clinical and radio-graphic follow-up evaluation after a minimum of 2 years. The mean follow-up period was 2.5 years. Five Hylamer liners have been revised for eccentric wear at a mean of 37 months postimplantation and one additional liner is scheduled to be revised at the 34-month follow-up evaluation (total, 4.2% failure rate due to wear). Survivorship of the Hylamer liner, with failure defined as revision due to eccentric wear, was calculated to be 86% (SE, 5.5%) at 4 years. The authors have discontinued the use of Hylamer acetabular liners because of this high early failure rate.
Medical subject headings
- Acetabulum
- Hip Prosthesis
Anatomy
- hip