Ultrasonography for diagnosis of rotator cuff tear.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 8741055.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In this study the clinical usefulness of ultrasonography in the diagnosis of rotator cuff tears was assessed. One hundred twenty-two patients undergoing preoperative ultrasonography were evaluated surgically. Three different ultrasonographic machines were applied with a 7.5-MHz linear-array transducer. Surgery confirmed a sensitivity of 79% for full-thickness tears and 50% for partial thickness tears. In 73 patients with negative ultrasonographic findings, the specificity was 100% for full-thickness tears and 90% for partial-thickness tears. The diagnostic sensitivity for full-thickness tears of the Hitachi EUB-340 system was significantly lower than those of Hitachi EUB-415 and EUB-515, and thus the former cannot be recommended for detection of rotator cuff tears. The sensitivity with regard to partial-thickness tears was also significantly lower with the Hitachi EUB-340 system as compared to the Hitachi EUB-515. There were no significant differences in false-positive rate among the three machines used. In conclusion, accurate ultrasonographic diagnosis of rotator cuff tear requires a high-resolution machine, as well as considerable experience in interpretation.