Pin removal in slipped capital femoral epiphysis: the unsuitability of titanium devices.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 8747353.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In an effort to predict which slipped capital femoral epiphysis fixation devices might cause fewer retrieval problems, 27 consecutive implant-retrieval procedures were reviewed, and mechanical tests using many currently available devices were performed. Problems occurred with both stainless steel and titanium devices if implanted for > 1 year. This tendency for increased problems was statistically significant for the titanium group, and analysis of a retrieved titanium screw showed evidence of osseointegration (direct bone-metal contact at > 90% of the interface) using backscattered image scanning electron microscopic analysis. These clinical data and data from mechanical testing suggest that stainless steel screws or trocar-tipped Steinmann pins are better suited for fixation of slipped capital femoral epiphysis.
Medical subject headings
- Bone Nails
- Epiphyses, Slipped
- Femur Head
- Stainless Steel
- Titanium
Anatomy
- femur