Superimposed inspiratory work of the Siemens Servo 300 ventilator during continuous positive airway pressure.

Tan, I K; Bhatt, S B; Tam, Y H; Buckley, T A; Oh, T E · Intensive Care Med · 1995

prospective_cohort · Level II

Where this comes from

Abstract

To compare the superimposed inspired work of breathing (SIW) of the Siemens Servo 300 ventilator with the Siemens Servo 900 C ventilator. Comparisons made at continuous positive airway pressure (CPAP) levels of 0, 4, and 8 cmH2O, and at trigger sensitivities of -1 and -2 cmH2O, and flow triggering. General intensive care unit in a University teaching hospital. 7 patients receiving CPAP. At all levels of CPAP, the SIW was significantly less with the Siemens Servo 300 ventilator as compared to the Siemens Servo 900 C ventilator despite similar trigger sensitivities. No significant difference was found in the SIW of the Servo 300 ventilator when comparing trigger sensitivities of -1 cmH2O, -2 cmH2O, and flow triggering. Different levels of CPAP had no effect on SIW. The Siemens Servo 300 ventilator entails less superimposed inspiratory work of breathing than the Siemens Servo 900 C ventilator.