Use of cemented all-polyethylene and metal-backed acetabular components in total hip arthroplasty. A comparative study.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 8776049.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The results of 115 primary total hip arthroplasties (97 patients) performed using the Charnley system and contemporary cementing techniques were reviewed. Fifty-five all-polyethylene cups and 60 metal-backed cups were used. The mean follow-up period was 84 months. No statistically significant difference was noted between groups for radiographic loosening and no cup has been revised to date. Survival analysis did not reveal a statistically significant difference between groups. Although all-polyethylene cups had significantly greater polyethylene thickness, no difference was noted for rate of wear. Increased loosening of the metal-backed cups was not observed. Polyethylene thickness is increased by the use of 22-mm heads and this may have accounted for the lack of difference in loosening rates of the all-polyethylene and metal-backed cups.
Medical subject headings
- Hip Prosthesis
Anatomy
- hip