Measurement of periprosthetic bone density in hip arthroplasty using dual-energy x-ray absorptiometry. Reproducibility of measurements.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 8792252.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To define the precision (reproducibility) of measurement of periprosthetic bone mineral density and bone mineral content, dual-energy x-ray absorptiometry scans were obtained on 45 randomly selected patients who had had a unilateral total hip arthroplasty within the previous 3 years. The coefficients of variation of the bone mineral density in the proximal Gruen zones were 5.0 and 5.3%, corresponding to errors of 0.07 and 0.11 g/cm2. The coefficients of variation of the bone mineral density for the distal zones averaged 2.8%, with an error of 0.08 g/cm2. The coefficients of variation of the bone mineral content were 4.8 and 2.9% for the proximal and distal zones. The contralateral femur was also scanned in 32 of the patients. For the contralateral femur bone mineral density, the coefficients of variation were 5.0% for the proximal zones and 4.8% for the distal zones. The bone mineral content was 6.0% for the contralateral regions. These results imply that differences in bone mineral density greater than 0.16 g/cm2 (2 standard errors) can be reliably measured. Dual-energy x-ray absorptiometry therefore provides a highly reproducible technique for quantitatively monitoring the changes in bone density that occur after total hip arthroplasty.
Medical subject headings
- Absorptiometry, Photon
- Bone Density
- Femur
- Hip Prosthesis
Anatomy
- femur
- hip