Transtracheal oxygen catheterization with pneumomediastinum and sudden death.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 8797438.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
A 69-year-old man with hypoxemic COPD underwent placement of a transtracheal oxygen (TTO) catheter. At 3 months, the catheter tract appeared mature with minimal erythema and no evidence of infection at the catheter site. The patient and his spouse were taught to remove and reinsert the catheter but were told to delay beginning the procedure due to erythema at the stoma site. Despite instructions not to remove the catheter for cleaning, the spouse removed the TTO catheter and attempted to reinsert it using the flexible metal cleaning rod. Subsequently, the patient suffered an acute episode of subcutaneous air and hemodynamic collapse resulting in death. Necropsy revealed a false catheter tract occluded by clotted blood and a defect in the platysma muscle where oxygen had dissected into the mediastinum. The patient died due to pneumomediastinum and cardiac tamponade.
Medical subject headings
- Cardiac Tamponade
- Catheterization
- Death, Sudden
- Lung Diseases, Obstructive
- Mediastinal Emphysema
- Oxygen Inhalation Therapy