Image quality and dose in spiral computed tomography.
Level III
Where this comes from
- Record sourced from PubMed, PMID 8798029.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Image quality and dose produced by a spiral CT with various pitch values were investigated on the basis of test-object images and measurements of CT dose indexes. No major difference in image quality or dose was observed when comparing sequential and spiral mode acquisition with a pitch of one. Increase in pitch value produces a loss of contrast which leads to a loss of detectability. Nevertheless, in realistic protocols the image quality in the transverse plane remains acceptable up to pitch values of 1.3-1.6.
Medical subject headings
- Radiation Dosage
- Radiographic Image Enhancement
- Tomography, X-Ray Computed