Influence of sample temperature on reflectance photometry and electrochemical glucometer measurements.
Where this comes from
- Record sourced from PubMed, PMID 8799638.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
A study was conducted to determine the influence of sample temperature on manual reflectance photometers, automatic reflectance photometers, and electrochemical glucometers. Aqueous and blood-based control solutions were tested at temperatures ranging from 25 to 44 degrees C. With the Accu-Chek 3, One Touch, and Satellite G glucometers, multiple glucose determinations were performed on each sample. The results indicate that the manual reflectance photometry glucometer is prominently influenced by variation in sample temperature. The effect of sample temperature is greatest at high glucose levels. Caution may be required in the interpretation of manual reflectance photometry glucometer measurements in febrile or hypothermic diabetic patients.