A proposed technique for intraoperative measurement of cervical spine stiffness.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 8805151.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Classically cervical biomechanical data have been obtained using cadaver models. Few methods exist for in vivo measurement of cervical biomechanics. A technique for intraoperative measurement of cervical motion segment stiffness (load-axial displacement) has been developed, enabling load/displacement data to be obtained during surgery at the spinal levels operated on and at adjacent cervical levels. The instrument used for measuring motion segment stiffness was a vertebral retractor equipped with displacement and strain transducers and designed to be placed over the shafts of Caspar distraction screws. Stiffness was measured under simulated anterior cervical surgical conditions. A preliminary study deciphering axial cervical spine load/displacement data on 12 cadavers was performed with this instrument. Using a distraction load of 88.8640 N, axial displacements of 2.110 +/- 0.258 mm (mean +/- standard error of the mean) at C3-C4, 2.367 +/- 0.304 mm at C4-C5, and 2.207 +/- 0.238 mm at C5-C6 were found. When data were evaluated for age, there was a significant decrease in stiffness with advancing age in two of the three spinal segments tested (P < 0.05 for C3-C4, P < 0.04 for C4-C5, P < 0.12 for C5-C6), which correlated with previous studies. These results suggest that this technique may be useful in assessing intraoperative motion-segment stability.
Medical subject headings
- Cervical Vertebrae
- Monitoring, Intraoperative