Frequency content of normal and diabetic plantar pressure profiles: implications for the selection of transducer sizes.
cross_sectional · Level IV
Where this comes from
- Record sourced from PubMed, PMID 8809630.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
How small do pressure transducers need to be in order to faithfully measure the plantar pressure profiles (PPPs) under normal and diabetic feet? In this study, pressures were collected from five diabetic and six non-diabetic subjects using a commercial measurement system with 25 mm2 transducers. Discrete Fourier Transform techniques were then used to determine (i) the spatial frequency content of diabetic and non-diabetic PPPs, and (ii) the effects of quadrupling the transducer area (from 5 mm x 5 mm to 10 mm x 10 mm). When the data were filtered to represent the effects of using 10 mm x 10 mm transducers, it was found that the ensuing reductions in peak pressure in the toe region (50 kPa) were significantly greater than in all other regions of the foot (p < 0.05). There was a significant correlation between pressure underestimations and measured peak pressures in the metatarsal regions. Based on data collected with 25 mm2 transducers it was concluded that transducer sizes greater than 6.36 mm x 6.18 mm (medio-lateral and antero-posterior directions) would result in sub-optimal sampling of PPPs.
Medical subject headings
- Diabetes Mellitus
- Foot