Mechanical failure of the intramedullary hip screw in a subtrochanteric femoral fracture.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 8814577.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The Gamma nail, an implant specifically designed for intertrochanteric and subtrochanteric femoral fractures, has been criticized for its high risk of secondary shaft fractures. A modified design, the intramedullary hip screw has recently been introduced to correct this complication. We present a case of mechanical failure of this new implant that occurred in a pathological subtrochanteric fracture. The centering sleeve of the implant became loose and migrated while the head screw penetrated the acetabulum. Refixation was successful.
Medical subject headings
- Bone Screws
- Femoral Fractures
- Fracture Fixation, Intramedullary
- Fractures, Spontaneous
Anatomy
- femur
- hip