High implant fracture incidence with Sutter silicone metacarpophalangeal joint arthroplasty.
retrospective_cohort · Level III
Where this comes from
- Record sourced from PubMed, PMID 8891979.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We performed a retrospective review of Sutter silicone metacarpophalangeal (MP) joint arthroplasties in 34 patients (42 hands, 168 implants) with rheumatoid arthritis. Patients were evaluated at an average of 27 months (minimum follow-up period, 12 months). Twenty percent of the implants were shown to be definitely fractured on final follow-up examination, and 45% followed for more than 3 years were definitely fractured. At the final follow-up examination, the average ulnar drift in intact implants was 11 degrees and in the fractured implants, 23 degrees. However, there was no correlation between implant fracture and patient satisfaction. Eighty percent of patients said they would undergo the procedure again. Because of a significantly higher implant fracture incidence at a relatively shorter follow-up period than that of most studies of silicone MP implants of the Swanson design, we have abandoned the use of the Sutter implant.
Medical subject headings
- Arthritis, Rheumatoid
- Joint Prosthesis
- Metacarpophalangeal Joint
- Silicones
Anatomy
- hand