Efficient real-time confocal microscopy with white light sources.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 8893003.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The main advantage of confocal microscopes over their conventional counterparts arises from their ability to optically 'section' nearly transparent materials; the thin image slices thus obtained can be used to reconstruct three-dimensional images, a capability which is particularly useful for the study of biological specimens. Confocal microscopes have previously used either a single laser-illuminated point-source and single point-detector (which are scanned in tandem across the object) or white-light illumination with multiple point-sources and detectors. Single-point-source systems, however, do not usually form images in real time and are restricted to using available laser wavelengths. Multiple-point-source systems, on the other hand, produce images in real time but use light very inefficiently--typically 1% or less is used for imaging. Here we demonstrate a white-light, multiple-point-source method which can in principle produce images in real time with light efficiencies as high as 50%. This system is likely to find broad practical application, particularly in the imaging of weakly reflecting or weakly fluorescent specimens.
Medical subject headings
- Microscopy, Confocal