Prevention of bacterial adherence to implant surfaces with a crosslinked albumin coating in vitro.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 8893783.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Titanium surfaces were coated with bovine serum albumin using carbodiimide, a crosslinking agent. The durability of the coated surfaces and the inhibitory effect of the albumin coating on bacterial adherence were tested in vitro for 20 consecutive days at 37 degrees C in phosphate buffered saline, with intermittent agitation. The results showed that only 10% of the coated bovine serum albumin decayed off the surface during the 20-day incubation period. The inhibition rate of the albumin coating on bacterial adherence remained high (greater than 8.5%) throughout the experiment. The results suggested potential use of this crosslinked albumin coating to reduce bacterial adherence and thus the subsequent possibility of prosthetic or implant infection in vivo.
Medical subject headings
- Bacterial Adhesion
- Cross-Linking Reagents
- Materials Testing
- Serum Albumin