Charge-burping theory correctly predicts optimal ratios of phase duration for biphasic defibrillation waveforms.
basic_science · Level V
Where this comes from
- Record sourced from PubMed, PMID 8901683.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
For biphasic waveforms, it is accepted that the ratio of the duration of phase 2 to the duration of phase 1 (phase-duration ratio) should be < or = 1. The charge-burping theory postulates that the beneficial effects of phase 2 are maximal when it completely removes the charge delivered by phase 1. It predicts that the phase-duration ratio should be < 1 when the time constant of the defibrillation system (tau s) exceeds the time constant of the cell membrane (tau m) but > 1 when tau s < tau m. This study tested the hypothesis that the optimal phase-duration ratio depends on tau s (the product of the defibrillator capacitance and pathway resistance). In a canine model of transvenous defibrillation (n = 8), we determined stored-energy defibrillation thresholds (DFTs) for biphasic waveforms from conventional capacitors (140 microF. tau s = 7.1 +/- 0.8 ms) and very small capacitors (40 microF. tau s = 2.0 +/- 0.2 ms). Each capacitance was tested with phase-duration ratios of 0.5, 1, 2, and 3. The duration of phase 1 approximated the optimal monophasic waveform, 6.3 +/- 0.7 ms for 140-microF waveforms and 2.8 +/- 0.2 ms for 40-microF waveforms. For 140-microF waveforms, the DFT was lower for phase-duration ratios < or = 1 than for phase-duration ratios > 1 (P = .0003). The reverse was true for 40-microF capacitors (P = .0008). There was a significant interaction between the effects of capacitance and phase-duration ratio on DFT (P = .0002). The lowest DFT for 40-microF waveforms was less than the lowest DFT for 140-microF waveforms (4.9 +/- 2.5 versus 6.4 +/- 2.4 J, P < .05). The optimal phase-duration ratio is < or = 1 for conventional capacitors and > 1 for small capacitors. This supports the predictions of the charge-burping theory.
Medical subject headings
- Electric Countershock