Total hip replacements using the ceramic Mittelmeier prosthesis.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 8913157.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
One hundred twelve consecutive primary total hip replacements were performed for osteoarthritis or osteonecrosis by a single surgeon using the Mittelmeier ceramic prosthesis without cement between January 1983 and July 1984. Ninety-three of these hips were followed prospectively to a minimum of 60 months and a median of 108 months. Five (5.4%) hips have been revised for loosening. Of the remaining hips, 77 (87.5%) were rated clinically satisfactory. Cumulative survival probability using revision as endpoint was estimated to be 92.7% at 5 years and 87.9% at 10 years. However, overall mechanical failure rates were 21.5% for the cup and 22.6% for the stem. There was no case of either femoral or acetabular osteolysis observed.
Medical subject headings
- Ceramics
- Femur Head Necrosis
- Hip Prosthesis
- Osteoarthritis
Anatomy
- hip
- femur