Two-dimensional coronary MR angiography without breath holding.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 8939224.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
To determine whether breath holding can be eliminated in two-dimensional magnetic resonance (MR) imaging of the coronary arteries by using real-time respiratory gating. Thirty-one subjects (20 healthy volunteers, 11 patients) underwent MR imaging. In 13 subjects, a respiratory monitoring belt was used, and in 18 subjects, a navigator echo was used. MR imaging was performed with breath holding, respiratory gating, and respiratory gating with two signals acquired. Three reviewers conducted a blinded review of the images, and overall image quality was rated on a scale from 1 (poor) to 5 (excellent). Respiratory gating with two signals acquired provided image quality superior to that with breath-hold imaging (3.7 vs 3.0, respectively; P < .05). Measurements of signal-to-noise ratio (14.5 for respiratory gating with two signals acquired and 11.9 for breath holding) supported the results of the image review. Navigator-echo gating provided better image quality than the monitoring belt (3.7 vs 3.1, respectively; P < .05). Breath holding may be eliminated by gating image acquisition to a real-time monitor of respiratory position. Respiratory gating enables improved resolution by means of acquisition of multiple signals, provides aligned sections of coronary arteries, and improves patient tolerance.
Medical subject headings
- Coronary Vessels
- Magnetic Resonance Angiography
- Respiration