Comparison of an optical catheter office arthroscope with a standard rigid rod-lens arthroscope in the evaluation of the knee.
rct · Level II
Where this comes from
- Record sourced from PubMed, PMID 8947405.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
In a prospective, randomized study, 47 patients underwent arthroscopic evaluation of the knee in an operating room setting with both a standard rod-lens arthroscope and a newer flexible optical catheter fiberoptic system. The goal of the study was to assess the diagnostic accuracy of the newer catheter system, which is recommended for use in the office setting. Forty-four patients were included in the data analysis, three were eliminated because we were unable to perform an adequate examination with the catheter scope secondary to intraarticular adhesions or excessive bleeding. A comparison of the two systems revealed an overall underestimation and underrecognition of intraarticular knee pathologic changes. Anterior cruciate ligament tears were missed in 3 of 21 knees; no posterior cruciate ligament tears were detected by the rod-lens arthroscope in 44 knees, but one false-positive result was "seen" using the optical catheter device. Similar trends in diagnosis (sensitivity, 25% to 67%; specificity, 96% to 99%) were seen for tears of the medial and lateral menisci, chondral lesions, and the identification of loose bodies. We conclude from the results of this study that the use of the optical catheter system for arthroscopic evaluation and treatment of the knee in the office setting may result in a significant compromise in visual acuity, resulting in missed and incorrect diagnoses.
Medical subject headings
- Arthroscopes
- Knee Joint
Anatomy
- knee
- tibia