Nonlinearity correction of brain perfusion SPECT based on permeability-surface area product model.

Tsuchida, T; Yonekura, Y; Nishizawa, S; Sadato, N; Tamaki, N; Fujita, T; Magata, Y; Konishi, J · J Nucl Med · 1996

case_control · Level III

Where this comes from

Abstract

It is well known that many cerebral perfusion tracers underestimate cerebral blood flow in high flow range. A model has been proposed to correct nonlinear relationship of flow and uptake of the tracers that accounts for the permeability-surface area product (PS model). We examined 43 patients in this study. To test the feasibility of this method for 123I-IMP, 99(m)Tc-HMPAO and 99(m)Tc-ECD, radioactivity ratios of cerebral regions to cerebellum (C/Cr) on SPECT images were compared with those of rCBF (F/Fr) measured by PET using the 15O CO2 steady-state method. Coefficient for correction in the PS model was estimated by the least squares method, and SPECT data were corrected using these coefficients. Estimated PS value by this method was highest in IMP (116 ml/min/100 g) followed by ECD (66 ml/min/100 g) and HMPAO (46 ml/min/100 g). The corrected SPECT data demonstrated an excellent linear relationship, which was close to unity, with rCBF. These results indicate that the PS model can be used for nonlinearity correction of brain perfusion SPECT.

Medical subject headings