Breath-hold MR imaging of focal liver lesions: comparison of fast and ultrasound techniques.
Level III
Where this comes from
- Record sourced from PubMed, PMID 8972320.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
The performance of breath-hold MR imaging using two T2-weighted hybrid sequences (TSE, TGSE), two T2-weighted single-shot sequences (HASTE, EPI-SE), and one T1-weighted gradient-echo sequence (FLASH) was compared with a standard conventional T2-weighted SE sequence in 20 patients with focal liver lesions. Liver signal-to-noise ratio was highest for the FLASH sequence (54.3 +/- 8.3) and the HASTE (41.1 +/- 12.5) sequence, whereas the highest spleen-liver contrast-to-noise ratio was obtained by the TSE sequence (38.9 +/- 20.7). Lesion-liver CNR was highest with the TSE sequence (63.9 +/- 21.4). With both TSE and HASTE significantly (p < 0.01) more lesions were detected as compared with SE and EPI-SE sequences. Our results indicate that breath-hold TSE and HASTE sequences will eventually replace conventional T2-weighted SE techniques, due to their insensitivity to motion artifacts, superior lesion detectability and inherently short acquisition times.
Medical subject headings
- Liver Diseases
- Liver Neoplasms
- Magnetic Resonance Imaging