Continuous end-tidal carbon dioxide monitoring during normofrequent jet-ventilation.
prospective_cohort · Level II
Where this comes from
- Record sourced from PubMed, PMID 8986189.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Continuous end-tidal carbon dioxide (ETCO2) monitoring during normofrequent jet-ventilation (NFJV) has not previously been successful and no correspondence between ETCO2 and arterial carbon dioxide (PaCO2) demonstrated. During NFJV jet-ventilation in 19 healthy volunteers, continuous ETCO2 was measured and compared to PaCO2 values. The original ETCO2 sampling line from our Datex CO2 measuring equipment was placed longitudinally against a suction catheter and both were wrapped with aluminium foil. They were placed 4-5 cm above the carina and the position was verified with a fiberoptic bronchoscope. The correlation between methods gave a Pearson's product moment correlation (r value) of 0.836, significant at the P < 0.001 level, indicating a close correlation between methods. A difference against mean scatter diagram confirmed that ETCO2 and PaCO2 measurements are of equal value in monitoring healthy patients during NFJV. A valuable method of continuous ETCO2 monitoring during NFJV is presented.
Medical subject headings
- Carbon Dioxide
- Monitoring, Intraoperative