Prosthesis displacement as a complication of reduction of a dislocated total hip arthroplasty.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 8986580.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Acute displacement at the cement-prosthesis interface of the femoral component of a hip arthroplasty that occurred as a consequence of attempted reduction of a dislocation is described. Although this complication is theoretically possible with a polished tapered stem design, it is previously unrecognized, and may be avoidable if fluoroscopic control is used during closed reduction of a dislocation.