Tracheal gas insufflation during pressure-control ventilation: effect of using a pressure relief valve.

Gowski, D T; Delgado, E; Miro, A M; Tasota, F J; Hoffman, L A; Pinsky, M R · Crit Care Med · 1997

basic_science · Level V

Where this comes from

Abstract

Pressure-control ventilation minimizes alveolar overdistention by limiting peak airway pressure, but a consequence of this pressure limitation may be a reduction in tidal volume with subsequent hypercarbia. Tracheal gas insufflation (TGI) can be used in combination with pressure-control ventilation to augment CO2 elimination. During pressure-control ventilation with continuous TGI, we observed that peak airway pressure increased above the set inspiratory pressure. Based on this observation, we investigated the ability of the pressure-control ventilator circuit to compensate for continuous TGI and the effect of insertion of a pressure relief valve to eliminate over-pressurization. University research laboratory. Using an artificial lung model, we studied the effects of continuous TGI with varying catheter flows (0, 2, 6, and 10 L/ min); ventilator frequencies (10 and 20 breaths/min); inspiratory duty cycles (0.33, 0.50, and 0.67); lung compliance (0.01, 0.02, and 0.04 L/cm H2O); and airway resistance (5, 20, and 50 cm H2O/L/sec) on: a) peak airway pressure; b) total inspiratory tidal volume; c) ventilator-derived tidal volume; and d) intrapulmonary pressure at end-exhalation (auto-PEEP). Tests were performed with and without a pressure relief valve whose threshold "pop-off" pressure was adjusted to match the set inspiratory pressure (35 cm H2O) for a total of 432 experimental conditions. Our data demonstrate that pressure-control ventilation augmented with continuous TGI can increase peak airway pressure above set inspiratory pressure due to delivery of a higher than intended tidal volume. Predisposing conditions include catheter flow rates of 6 and 10 L/min, long inspiratory time, low compliance, and low resistance. With the pressure relief valve, peak airway pressure was maintained at the set inspiratory pressure and total inspiratory tidal volume remained constant. A pressure relief valve is a necessary adjunct to maintain peak airway pressure at set inspiratory pressure and keep total inspiratory tidal volume constant when continuous TGI is administered in conjunction with pressure-control ventilation.

Medical subject headings