Mechanical evaluation of four- versus eight-pin halo fixation.
biomechanical · Level V
Where this comes from
- Record sourced from PubMed, PMID 8989714.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Fixation of the halo device in the skulls of young children is sometimes compromised because of a thinner cortex and the increased jarring and trauma that small children experience during their everyday activities of life at that age. To assess the increased fixation that an eight-pin halo fixation would provide over the more standard four-pin fixation, the stiffness of fixation of a halo orthosis was evaluated in four dried human cadaver skulls. Two different fixation systems were used: the first one with four pins in the standard locations and the second with four additional pins. A compressive force was applied to the halo ring at a constant rate of 1.27 mm/min. Load and deflection data were recorded simultaneously until a force of 400 N was reached. Results for the two-pin systems were analyzed and compared statistically by using paired t tests. Statistically significant differences were established between the two sets of results with the 8-pin system providing an increase of 24.2% in the stiffness of fixation.
Medical subject headings
- Skull
- Traction