Narrow photon beam dosimetry for linear accelerator radiosurgery.

Gotoh, S; Ochi, M; Hayashi, N; Matsushima, S; Uchida, T; Obata, S; Minami, K; Hayashi, K et al. · Radiother Oncol · 1996

basic_science · Level V

Where this comes from

Abstract

The dosimetric characteristics of linear accelerator radiosurgery for 10-MV X-ray were measured. Measurement of the relative output factor and tissue maximum ratio with a microchamber produced results equivalent to those of measurement with X-ray film. The 80% isodose level width measured with the microchamber was significantly smaller than that measured with the X-ray film. For the measurement of relative output factor and tissue maximum ratio, a microchamber seems to be the more appropriate choice. X-Ray film was found to be suitable for beam profile measurement.