Three-dimensional polyethylene wear of a press-fit titanium prosthesis. Factors influencing generation of polyethylene debris.

Devane, P A; Horne, J G; Martin, K; Coldham, G; Krause, B · J Arthroplasty · 1997

retrospective_cohort · Level III

Where this comes from

Abstract

Between 1985 and 1990, 108 consecutive Harris-Galante I (Zimmer, Warsaw, IN) total hip arthroplasties were performed by four surgeons at a single hospital. At the time of follow-up evaluation, 80 hips were available for review. The mean rate of linear wear was 0.15 mm/y, mean rate of three-dimensional femoral head displacement was 0.21 mm/y, and mean rate of volumetric wear was 121 mm3/y. Eight hips (10%) in this series had radiologic osteolysis around either the femoral or acetabular component. A significantly greater volumetric wear rate was found in patients who were younger, those with a higher activity level, those who received a 32-mm-diameter femoral head, and those with vertical orientation of their acetabular component. No relationship could be made with patient weight, gender, Harris hip score, or cup diameter.

Medical subject headings

Anatomy