Heterophilic antibodies produce spuriously elevated concentrations of the MB isoenzyme of creatine kinase in a selected patient population.
Level II
Where this comes from
- Record sourced from PubMed, PMID 9128261.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
Dual-site murine antibody-based immunoassays are commonly used in clinical laboratories to quantitate the MB isoenzyme of creatine kinase (CK-MB). Because the serum level of CK-MB is a relatively specific and sensitive indicator of myocardial ischemic damage, accurate quantitation is essential for a correct diagnosis. Heterophile antibodies (eg, human anti-murine antibodies) can interfere with these assays, however, and produce erroneous results. A subpopulation of 19 surgical patients with colorectal carcinoma who had received injections of an 125I-labeled murine monoclonal antibody directed against a tumor-associated glycoprotein was studied. Serum specimens from eight patients (42%) showed a marked increase in the level of CK-MB and normal total CK concentrations. The increased concentrations of CK-MB, which were attributed to interference by human antimurine antibodies, were substantially reduced in these specimens after a heterophile blocking reagent was added. However, this reagent did not significantly alter the serum level of CK-MB in patients who had clinical evidence of acute myocardial ischemia.
Medical subject headings
- Antibodies, Heterophile
- Antigens, Neoplasm
- Carcinoma
- Colorectal Neoplasms
- Creatine Kinase
- Glycoproteins