Stent deformation and intimal hyperplasia complicating treatment of a post-carotid endarterectomy intimal flap with a Palmaz stent.
case_report · Level V
Where this comes from
- Record sourced from PubMed, PMID 9129637.
- No licence information is recorded for this record.
- Because redistribution is not established, this page shows the abstract only. Follow the links below for the full text.
Abstract
We report a case of common carotid artery Palmaz stent placement for treatment of an intimal flap after surgical endarterectomy. Despite technical success with an excellent immediate result, a significant stenosis detected by duplex sonographic examination developed at 10 months. This stenosis, the result of stent compression and intimal hyperplasia, illustrates the previously theoretic risk associated with placement of the balloon-expandable stent in a compressible site such as the cervical carotid artery. In addition, we demonstrate that significant intimal hyperplasia may occur after carotid artery stent placement, potentially limiting long-term patency.
Medical subject headings
- Endarterectomy, Carotid
- Stents
- Tunica Intima