Occupational exposure to desflurane and isoflurane during cardiopulmonary bypass: is the gas outlet of the membrane oxygenator an operating theatre pollution hazard?

Hoerauf, K; Harth, M; Wild, K; Hobbhahn, J · Br J Anaesth · 1997

rct · Level II

Where this comes from

Abstract

We have compared occupational exposure to isoflurane and desflurane during cardiopulmonary bypass, with and without a scavenging system at the membrane oxygenator outlet. Trace concentrations of volatile anaesthetics were measured by a direct reading instrument in 40 elective heart surgery procedures. Measurements were obtained in the breathing zones of the anaesthetist and perfusionist. When a scavenging system was used, median desflurane values were less than 0.3 ppm and isoflurane values less than 0.2 ppm. Without a scavenging system values were, in general, three- (isoflurane) to five- (desflurane) fold higher. We conclude that the use of a scavenging system at the membrane oxygenator outlet can reduce occupational exposure to volatile anaesthetics. We therefore recommend routine use of scavenging devices during cardiopulmonary bypass.